ALPHA ACTIVE® 4 SYSTEM Service Manual

Alpha Active 4 Service Manual

CONTENTS CHAPTER 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Numbering and Cross-Referencing in this Manual . . . . . . . . . . . . . . . . . . . . . 1 Warnings, Cautions and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Technical Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Alpha Active 4 Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Cycle Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Alpha Active 4 Mattress Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Pneumatic and Electrical Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Controls and Indicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 On/Off Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Alarm/Mute . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Mode Select . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Auto-firm Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Pressure Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Cycle Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Alarms and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Service Indicator/Pump Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Power Fail Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

CHAPTER 2 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Troubleshooting Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

CHAPTER 3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Pump Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Pump Maintenance Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Mattress Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Mattress Maintenance Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Alarm Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Power Fail Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Low Pressure Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Service Pump/Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

Issue 1 April 2010

SER0018

Page i

Alpha Active 4 Service Manual

CHAPTER 4 Testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Mattress Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Pump Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Pump Flow, Pressure, and Function Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Flow and Pressure Checks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Electrical Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Electrical Safety Checks - Class II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Functional Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Power Fail Alarm Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Low Pressure Alarm Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Service Alarm Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Mattress Inflation Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 End Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

CHAPTER 5 Pump Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Removing the Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Replacing the Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Removing The Bed Hooks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Replacing the Bed Hooks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Removing the Mains/Power Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Installing the Mains/Power Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Removing the Front Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Replacing the Front Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Removing the Display PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Installing the Display PCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Removing the Compressor Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Installing the Compressor Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Removing the Microswitch Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Installing the Microswitch Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Removing the Rotor/Stator Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Installing the Rotor/Stator Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Removing the Gearbox Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Installing the Gearbox Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Removing the Transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Installing the Transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Removing the Main PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Page ii

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual Replacing the Main PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Removing Power Fail Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Replacing Power Fail Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Removing the On/Off Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Installing the On/Off Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Removing the Fuse Holder Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Installing the Fuse Holder Housing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Removing the Mains/Power Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Installing the Mains/Power Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Removing the Terminal Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Installing the Terminal Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

CHAPTER 6 Mattress Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Alpha Active 4 Cell Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Removing The Top Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Installing the Top Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Removing Cell Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Installing Cell Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Replacing Manifold Tubing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Removing CPR Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Installing CPR Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Removing Static Feed Check Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Installing Static Feed Check Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Removing 3-Way Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Installing 3-Way Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

CHAPTER 7 Technical Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Mattress . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Mattress Size Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Cover Specification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Cleaning Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

CHAPTER 8 Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Pump Assembly Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Issue 1 April 2010

SER0018

Page iii

Alpha Active 4 Service Manual Mattress Assembly Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

CHAPTER 9 Service Contact Details. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Page iv

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Introduction

CHAPTER 1 INTRODUCTION 1

About This Manual

ArjoHuntleigh strongly recommend that their equipment is only serviced by trained personnel and provide courses for customers who wish to become licensed to service their own equipment. In no event will ArjoHuntleigh be responsible for any service performed by customers or third parties. This manual contains information on maintenance, servicing, repair, troubleshooting and testing for the Alpha Active 4 system, which comprises of a pump and a mattress replacement. Read and understand this manual before attempting to service or repair the equipment.

Numbering and Cross-Referencing in this Manual For all chapters in this manual: •

Page, section and paragraph numbering re-start at “1”.



Figure and table numbering continue from the previous chapter.



Cross-references, which include a chapter number (and/or chapter title) refer to text in a different chapter. Cross-references which do NOT include any chapter number (or chapter title) refer to text within the same chapter.

Warnings, Cautions and Notes WARNINGS given in this manual identify possible hazards in procedures or conditions which, if not correctly followed, could result in death, injury or other serious adverse reactions. CAUTIONS given in this manual identify possible hazards in procedures or conditions which, if not correctly followed, could result in equipment failure or damage. Notes given in this manual are used to explain or amplify a procedure or condition.

WARNING: BEFORE PERFORMING ANY SERVICE OR MAINTENANCE PROCEDURES, ENSURE THAT THE EQUIPMENT HAS BEEN ADEQUATELY DECONTAMINATED. WARNING: VOLTAGES IN EXCESS OF 30 VOLTS RMS OR 50 VOLTS DC CAN, IN CERTAIN CIRCUMSTANCES, BE LETHAL. WHEN WORKING ON EQUIPMENT REQUIRING EXPOSURE TO LIVE, UNPROTECTED CONDUCTORS WHERE SUCH VOLTAGES ARE PRESENT, EXTREME CARE MUST BE EXERCISED.

2

Technical Description

The Alpha Active 4 system comprises of a mattress replacement and pump. The product can be used on hospital and domestic beds in both domestic and care environments. The bed base must be suitable for the specified mattress size.

Alpha Active 4 Pump The Alpha Active 4 pump comprises of a moulded case with non slip feet on the base and integral hanging brackets.

Issue 1 April 2010

SER0018

Chap 1 Page 1

Introduction

Alpha Active 4 Service Manual

The controls are situated on the front of the pump and a sophisticated alarm system differentiates between normal operation and genuine system faults. If an alarm situation is detected an indicator will illuminate on the front of the pump and an audible warning will sound.

Cycle Options The pump has a user selectable option for operating on either a ten minute, or twenty minute cycle. 10 minutes Cycle In this alternating mode, the system has a cycle time of 10 minutes. Each full cycle consists of 4.5 minutes inflate, 30 seconds change over, 4.5 minutes deflate and 30 seconds changeover. 20 minutes Cycle In this alternating mode, the system has a cycle time of 20 minutes. Each full cycle consists of 9.5 minutes inflate, 30 seconds change over, 9.5 minutes deflate and 30 seconds changeover.

Alpha Active 4 Mattress Replacement Each of the air cells is split into two sections. The base section of each being linked together to provide a static section which provides protection from the effects of bottoming out. The Alpha Active 4 cover is water resistant and vapour permeable to enhance patient comfort whilst protecting the cells from ingress of contaminants. The cover is simple to clean in situ, but may easily be removed for laundering, preventing cross contamination. In the event of cardiac arrest the mattress replacement can be rapidly deflated to allow cardiac resuscitation procedure (CPR) to be performed. Table 1 - Available Mattress Sizes Part No.

Description

Spare Cover

Length mm

Width mm

Height mm

648322

ALPHA ACTIVE 4 MR 90

648437

1950 (77")

900(35")

200 (8")

648324

ALPHA ACTIVE 4 MR 85

648461

1950 (77")

850(33")

200 (8")

648326

ALPHA ACTIVE 4 MR 85 PU

648461

1950 (77")

850(33")

200 (8")

Chap 1 Page 2

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

3

Introduction

Pneumatic and Electrical Diagrams

Figure 1 - Pneumatic System Diagram

Issue 1 April 2010

SER0018

Chap 1 Page 3

Introduction

Alpha Active 4 Service Manual

*Connectors J1, J2 and J7 must be connected to their relevant terminals on the Main PCB. Figure 2 - Electrical Diagram

Chap 1 Page 4

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Introduction

4

Front Panel

4.1

Below is a diagram of the Alpha Active 4 pump front panel. For more detail on the function of each button, see below. Cycle Time

Alarm Mute

Comfort Control

5

Mode Selector

Auto-firm Mode

Alarms

Controls and Indicators On/Off Switch Located on the right side of the pump unit, the On/Off switch will illuminate green when the pump is on.

Alarm/Mute During an alarm condition the sound of the alarm can be muted by pressing this button. The yellow indicator will then flash intermittently but the alarm will be muted.

Mode Select The mode selector allows the mode of operation to be selected. Two modes of operation are available, Alternating or Static (non-alternating) The indicator next to the respective icon indicates the currently selected mode.

.

Auto-firm Mode Selects the Auto-firm, non-alternating mode. The orange indicator will illuminate when the pump is in this mode. Autofirm operates for 30 minutes.

Pressure Control Two buttons (+ and -) are used to set the pressure inside the mattress replacement and therefore the amount of support the patient receives. The pressure setting is indicated by the green indicator above the selected setting.

Cycle Time A single button allows the cycle time to be selected for the most appropriate patient therapy. The cycle time setting is indicated by the green indicator above the selected setting.

Issue 1 April 2010

SER0018

Chap 1 Page 5

Introduction

6

Alpha Active 4 Service Manual

Alarms and Indicators Low Pressure Indicator The red Low Pressure indicator is illuminated whenever the pump detects low pressure within the mattress replacement. An audible alarm will sound unless cancelled by the mute button. The indicator will extinguish once normal pressure is reached. Refer To “Troubleshooting” on page 14 for possible causes of Low Pressure.

Service Indicator/Pump Fault The yellow Service/pump fault indicator will illuminate and remain on if the pump has detected an internal fault. An audible alarm will sound unless cancelled by the mute button.

Power Fail Indicator The red Power Fail indicator will illuminate when a mains power failure has been detected. An audible alarm will sound until power is restored or the pump is switched off using the on/off switch.

Chap 1 Page 6

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Troubleshooting

CHAPTER 2 TROUBLESHOOTING 1

Troubleshooting Table

The following table contains fault symptoms, their possible causes and suggests steps to rectify the problem. These may occur during initial installation and power up. Table 2 - Troubleshooting Table Symptom Low Pressure Alarm (Mattress)

Low Pressure Alarm (Pump)

Power Fail Alarm

Service Alarm

Issue 1 April 2010

Possible Cause

Action

Tube-set not connected securely

Check the tube-set connector and ensure it is securely fitted to the pump.

CPR not fully closed.

Close CPR Unit

Cell Leak.

Trace and replace faulty part.

Damaged/Disconnected tube.

Check/Replace.

Faulty Compressor

Check/Replace. (Chapter 5, Page 13, Section 12).

Blocked Air Filter

Check/Replace. (Chapter 5, Page 3, Section 2).

Power has been removed from the pump.

Re-apply power or switch the pump off.

Blown Fuse.

Check/Replace. (Chapter 5, Page 7, Section 6).

Pump has detected an internal fault.

Trace and replace faulty part.

Faulty Microswitch

Test and Replace if required. (Chapter 5, Page 15, Section 14).

Faulty Gearbox

Check/Replace if required. (Chapter 5, Page 15, Section 18).

SER0018

Chap 2 Page 1

Troubleshooting

Chap 2 Page 2

Alpha Active 4 Service Manual

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Maintenance

CHAPTER 3 MAINTENANCE WARNING: BEFORE DISMANTLING THE PUMP, ENSURE UNIT HAS BEEN ISOLATED FROM THE POWER SUPPLY BY REMOVING THE CORD PLUG FROM THE WALL OUTLET. WARNING: A POTENTIAL ELECTRICAL SHOCK HAZARD EXISTS WHEN THE BACK COVER IS REMOVED, EVEN WITH THE PUMP SWITCHED OFF. CAUTION: Static Sensitive Devices This pump should only be opened by personnel trained in ESD methods and with appropriate equipment and anti-static protection.

1

Pump Service

A pump should be serviced every 12 months. To carry out a service on the pump, do the following: 1.1

Carry out the Pump Maintenance Checks (Refer to Page 1, Section 2).

1.2

Replace the air inlet filter.

2

Pump Maintenance Checks

If any parts are found to be damaged they must be replaced in accordance with Pump Repair. 2.1

Visually inspect the following for damage, wear and potential faults: • Base of Pump

Check base is free from debris.

• Inlet Filter

Check condition.

• Compressor

Check compressor seating.

• Electrical Connections

Check connections are secure and completely insulated.

• Tubing

Check completely pushed on and are kink free.

• Power Switch

Check switch illuminates when ON.

• Wiring

Check wiring is intact. Check electrical insulation are not damaged.

• Air flow

Check air flow is alternating.

• Outlet Tubes

Check tubes are secure.

Issue 1 April 2010

SER0018

Chap 3 Page 1

Maintenance

3

Alpha Active 4 Service Manual

Mattress Service

To carry out a service on the mattress, do the following: 3.1

Carry out the Mattress Maintenance Checks (Refer to Page 2, Section 4).

4

Mattress Maintenance Checks

If any parts are found to be damaged they must be replaced in accordance with Mattress Repair. 4.1

Visually inspect the following for damage, wear and potential faults: • Base and Top Cover

Check for tears, staining, clarity of printed labels.

• Base/Top cover Zip

Check the zip runs freely and condition of zip teeth.

• Fixing Straps

Check condition and security.

• Connectors

Check for damage and security.

• Tubing

Check for security and ensure no kinks or twists.

• Cells

Check for damage.

• Manifold

Check for damage and security.

• Press Studs

Check condition.

• CPR

Check operation of Fast Deflate mechanism.

• Tubeset Assembly

Check for damage and security of connection.

4.2

Carry out a function test on the pump/mattress system in accordance with Chapter 4.

5

Alarm Functions

Power Fail Alarm Switch pump off at mains/power socket, leaving the pump switched on. The red power fail indicator will illuminate when a mains power failure has been detected. An audible alarm will sound until power is resumed or the pump is switched off using the on/off button.

Low Pressure Indicator 5.1

Block the air outlets on the pump.

5.2

Switch the pump on and allow Low Pressure indicator to extinguish.

5.3

Remove the blockage from the air outlets. The red Low Pressure indicator will illuminate and and audible alarm will sound.

Service Pump/Fault The yellow service/pump fault indicator will illuminate and remain on if the pump has detected an internal fault. Refer to Troubleshooting to help diagnose the fault, refer to Chapter 2.

Chap 3 Page 2

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Testing

CHAPTER 4 TESTING CAUTION: Electrostatic discharge can seriously damage the Main PCB. Prior to replacing it, adequate earthing precautions must be taken. Note: In all cases the pump and mattress must be tested together as a system.

1

Mattress Visual Inspection

Ensure that the mattress is checked as follows: 1.1

All studs must be secure and all inlet connectors fitted securely.

1.2

The mattress contains 19 cells which are studded to both sides of the mattress. The bottom of each cell connected to the static feed, the top half connected to the alternating feed (the head section cells top half are connected to the static feed). Ensure that the cells are retained centrally by the loops in the base cover.

2

Pump Visual Inspection

2.1

Ensure there are no cracks or marks on the pump case. Ensure the labels are intact and readable. Check there are no marks or fractures in the Pump outlet module.

2.2

Ensure the mains lead is fully secure and there are no cuts or damage to the cable.

3

Pump Flow, Pressure, and Function Test

You will also need: •

A sphygmomanometer or pressure gauge (0-160mmHg)



A flow meter (0-10 litres/minute)

To carry out a system functional test complete the following procedures: •

Flow and Pressure Checks



Electrical Testing



Functional Test

The following procedures must be carried out after a service or major repair of the pump. To test/ calibrate pump correctly, the following equipment (or equivalent) is required: Note: All test equipment must be calibrated to national or international standards.

Flow and Pressure Checks 3.1

Set the pressure control on the pump to maximum (200kg) setting.

3.2

Connect a dual flow pressure meter, via a valve, to whichever outlet is working.

3.3

Open the valve fully. The free flow rate should not be less than specified in the table below. Air Flow @ 20mmHg 6 litres/min

Issue 1 April 2010

Air Flow @ 50mmHg 3 litres/min

SER0018

Chap 4 Page 1

Testing 3.4

Alpha Active 4 Service Manual

Close the valve and operate the pressure control on the pump from minimum (20kg) to maximum (200kg) setting. The pressure indicated on the gauge (in mmHg) should be as specified in the table below. Minimum Pressure @ 20kg

Maximum Pressure @ 200kg

25mmHg ± 5mmHg

75mmHg ± 5mmHg

3.5

Repeat for the remaining ports.

4

Electrical Testing

To test and calibrate the pump unit correctly, the following equipment is required: Item

Test Equipment

1

Insulation Resistance Tester (Megger), 500 Vdc

2

Portable Appliance Tester

3

Multimeter / Continuity Tester

4

Dielectric strength tester (Flash tester) 3.0 kVac with current limit (CHECK 4.0kVac??)

Electrical Safety Checks - Class II There are several electrical safety checks that must be carried out after breakdown repairs, rental rechecks and servicing. Where alternatives are given, the test will depend upon the available equipment. The tests are as follows: •

Insulation Resistance Test (Megger Test) or



Dielectric Strength (Flash Test)

Insulation Resistance Test This test checks the integrity of the appliance’s insulation. For all appliances, this test is applied between the connected Live and Neutral wires and the appliance’s metal compressor assembly box. Specification: 500 Vdc is applied to the insulation and the measured resistance must be greater than 2MΩ. This test can be conducted by using a Portable Appliiance Tester. Also check the continuity of the mains cable.

Dielectric Strength Test (Flash Test) This test shows the response of the insulation to high ac-voltage stress, to the efforts of the capacitive current and gives an early warning of any electrical problems developing in the appliance.

WARNING: DANGER OF ELECTRIC SHOCK. DO NOT TOUCH ANY EXPOSED PARTS WHILE CONDUCTING THIS TEST. DO NOT TOUCH ANY PART OF YOUR BODY WITH THE CONTACTS OF THE PROBES. WARNING: THIS EQUIPMENT SHOULD NOT BE USED IF YOU HAVE A HEARING AID OR PACEMAKER FITTED, DUE TO THE POSSIBILITY OF ELECTROMAGNETIC DISTURBANCE. Specification: 4.0 kVac is applied between the connected Live and Neutral wires and the foil applied to the device's enclosure. No breakdown should occur.

Chap 4 Page 2

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Testing

Note: The voltage levels used for this test, may stress and weaken the insulation. This test is therefore not recommended as a routine test. It should only be used after a major assembly/disassembly has been carried out.

Functional Test Tests must be carried out using a 220-240V 50Hz power supply. Test Setup and Initial Switch On 4.1

Set the pressure control to minimum (20kg) and ensure that all three outlet ports are open to atmosphere.

4.2

Connect the mains/power cable to a suitable power socket outlet and then set the on/off switch to ‘on’. Verify that the green indicator in the switch is illuminated. Verify that the red LOW PRESSURE indicator is illuminated. Use the fingers or flowtron stopper (part number CAS077) to close all three ports and verify that the LOW PRESSURE indicator goes off.

Power Fail Alarm Check 4.3

With the pump operating, disconnect the power supply at the power socket outlet. Verify that, within 5 seconds, the red POWER FAIL indicator flashes in conjunction with an audible alarm. Reconnect the power supply. Verify the POWER FAIL indicator is extinguished.

Low Pressure Alarm Check 4.4

Close all three ports. Operate the on/off switch to ‘on’. verify that, after 10 seconds, the SOLID LOW PRESSURE indicator is extinguished.

4.5

Open all ports to atmosphere. Verify that, after 10 seconds, the red LOW PRESSURE indicator flashes in conjunction with the audible alarm.

Service Alarm Check To check the operation of the Service alarm it is necessary to have the case halves open.

WARNING: ENSURE PUMP IS DISCONNECTED FROM THE POWER SUPPLY. 4.6

Open the case halves (Refer to Chapter 5, Page 9, Section 8).

4.7

Disconnect the micro switch from the PCB connector, J4 (Chapter 1, Page 4, Section 2).

4.8

Close the case halves.

4.9

Connect the mains/power cable to a suitable power socket outlet and close all three ports. Operate the on/off switch to ‘on’.

4.10

The service indicator should operate within 10 seconds.

4.11

Operate the on/off switch to ‘off’ position, disconnect the power supply at the power socket outlet.

4.12

Open the case halves and reconnect the micro switch to the PCB connector, J4. (Chapter 1, Page 4, Section 2).

4.13

Close the case halves (Refer to Chapter 5, Page 9, Section 8).

4.14

Connect the mains/power cable to a suitable power socket outlet and close both ports. Operate the on/off switch to ‘on’.

4.15

The service indicator should be extinguished.

5

Mattress Inflation Test

5.1

Test the complete system to ensure correct operation.

Issue 1 April 2010

SER0018

Chap 4 Page 3

Testing

Alpha Active 4 Service Manual

6

End Setting

6.1

Disconnect the mattress from the pump and coil up the mains cable. Expel all the air from the mattress. The system is now ready to be packed.

Chap 4 Page 4

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Pump Repair

CHAPTER 5 PUMP REPAIR WARNING: BEFORE DISMANTLING THE PUMP, ENSURE UNIT HAS BEEN ISOLATED FROM THE POWER SUPPLY BY REMOVING THE CORD PLUG FROM THE WALL OUTLET.

1

General

This chapter details repair procedures for the Alpha Active 4 Pump Unit. All repairs should be carried out by ArjoHuntleigh approved service personnel. If any cable ties are cut so as to enable removal of parts, replacement ties must be fitted as part of the install procedure. Table 3 defines the test requirements which must be carried out following certain repairs: To carry out a flow, pressure and function test on the pump refer to Chapter 4, Page 1, Section 3. To carry out the electrical tests on the pump refer to Chapter 4, Page 2, Section 4.

Table 3 - Repair to Testing Requirements Components / Assemblies

Function Test

Electrical Safety Tests

Front Casing

Yes

Yes

Rear Casing

No

Yes

Display PCB

Yes

Yes

Main PCB

Yes

Yes

Bed Hooks

No

Yes

Compressor

Yes

Yes

Rotor/Stator Assembly

Yes

Yes

Mains Lead

No

Yes

Mains Switch

No

Yes

Terminal Block

No

Yes

Transformer

No

Yes

Air Filter

Yes

Yes

Oulet Connector

Yes

Yes

Issue 1 April 2010

SER0018

Chap 5 Page 1

Pump Repair

Alpha Active 4 Service Manual

10

20

Figure 3 - Replace the Air Filter Table 4 - Replace the Air Filter Parts List Item

Part Number

10

648402

Rear Case Alpha Active 3 and 4

1

20

648405

Air Filter Cover

1

-

648404

Air Filter Replacement

1

Chap 5 Page 2

Description

SER0018

Qty

Issue 1 April 2010

Alpha Active 4 Service Manual

Pump Repair

2

Removing the Air Filter

2.1

Place the pump unit face down on a clean, flat surface.

2.2

Press firmly on the retaining clip of the air filter housing, and lift the Air Filter housing (Fig 3, Item 20) away.

2.3

Carefully lever out the the Air Filter from the housing (Fig 3, Item 20).

3

Replacing the Air Filter

3.1

Place the new air filter in the air filter housing (Fig 3, Item 20).

3.2

Place the air filter housing (Fig 3, Item 20) into the rear case (Fig 3, Item 10) by ensuring that the two securing lugs align with the holes provided.

3.3

Carefully press down on the air filter housing retaining clip and ensure it the housing has been securely fitted into the rear case (you should hear a ‘CLICK’ confirming proper fitment).

Issue 1 April 2010

SER0018

Chap 5 Page 3

Pump Repair

Alpha Active 4 Service Manual

30 10

20

Figure 4 - Replacement of Bed Hooks Table 5 - Bed Hook Parts List Item

Part Number

10

648402

20

REF

30

648403

Chap 5 Page 4

Description

Qty

Rear Case Alpha Active 3 and 4

1

Bed Hook Cover

2

Bed Hook Alpha Active 3 and 4

2

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Pump Repair

4

Removing The Bed Hooks

4.1

Place the pump unit face down on a clean, flat surface.

4.2

Ensure the Bed Hook is folded against the pump.

4.3

Using a small flat tool, remove the Bed Hook Cover by inserting the tool into the aperture on the top edge of the cover.

4.4

Remove the two securing screws from the Bed Hook Bracket. Note: You may be required to gently lever the bracket away from the pump in order to fully remove it. If so, do this via the top edge of the recess, and ensure you place a free hand over the bracket to prevent its loss.

4.5

Rotate the hook 90° and remove from the pump casing.

4.6

Repeat Para 4.2 to Para 4.5 for the other Hook.

5

Replacing the Bed Hooks

5.1

Firmly insert the new Hook (Fig 4, Item 30) at 90° to the pump case, into the corresponding recess.

5.2

Whilst applying pressure to the hook, rotate it so it is folded in against the pump.

5.3

Position the Bed Hook Bracket and fit the two retaining screws.

5.4

To refit the Bed Hook Cover. 5.4.1

Clean all previous glue residue off from both the cover and the recess lip.

5.4.2

Apply an instant adhesive (Loctite 460 or similar) to the lip within the recess of the pump case.

5.4.3

Firmly press the Bed Hook cover into position.

5.5

Repeat for the other Hook.

5.6

Assemble the case (Refer to Page 9, Section 9).

Issue 1 April 2010

SER0018

Chap 5 Page 5

Pump Repair

Alpha Active 4 Service Manual

10

20 Figure 5 - Position of Mains/Power Fuses Table 6 - Mains/Power Fuses Parts List Item

Part Number

10

648402

Rear Case

1

20

MIS106

Mains Fuse (1A Time Lag T Fuse)

2

Chap 5 Page 6

Description

SER0018

Qty

Issue 1 April 2010

Alpha Active 4 Service Manual

6

Pump Repair

Removing the Mains/Power Fuses

WARNING: BEFORE DISMANTLING THE PUMP, ENSURE UNIT HAS BEEN ISOLATED FROM THE POWER SUPPLY BY REMOVING THE CORD PLUG FROM THE WALL OUTLET. 6.1

Place the pump unit face down on a clean, flat surface.

6.2

Using a flatbladed screwdriver, remove the fuse holder from the back of the rear case by unscrewing anti-clockwise.

6.3

Pull the blown fuse from its holder.

7

Installing the Mains/Power Fuses

7.1

Firmly push the new fuse into its holder.

7.2

Place the fuse holder into position on the back of the rear case.

7.3

Using a flatbladed screwdriver tighten clockwise until tight.

Issue 1 April 2010

SER0018

Chap 5 Page 7

Pump Repair

Alpha Active 4 Service Manual

30

30

60

20

30

10

30

40

50

Figure 6 - Front Case Removal/Fitment Table 7 - Case Assembly Item

Part Number

10

648401

Front Case Alpha Active 4

1

20

648402

Rear Case Alpha Active 3 and 4

1

30

REF

Screw (M3x15mm Pozi Pan Head, PT Type)

5

40

648420

Display PCB Alpha Active 4

1

50

REF

Main PCB - Display PCB Ribbon Connector

1

60

648423

Hose Connector - Alpha Active 4

1

Chap 5 Page 8

Description

SER0018

Qty

Issue 1 April 2010

Alpha Active 4 Service Manual

8

Pump Repair

Removing the Front Case

WARNING: BEFORE DISMANTLING THE PUMP, ENSURE UNIT HAS BEEN ISOLATED FROM THE POWER SUPPLY BY REMOVING THE CORD PLUG FROM THE WALL OUTLET. 8.1

Place the pump unit face down on a clean, flat surface.

8.2

Unscrew the five screws (Fig 6, Item 30).

8.3

Carefully turn the pump over, collecting the five screws (Fig 6, Item 30), keeping them in a safe place.

8.4

Carefully pull apart the two cases, ensuring the compressor remains in the rear case. Note: The front case contains a PCB connected via a ribbon to the main PCB in the rear case.

8.5

Within the front case, on the Display PCB pull apart the two ribbon securing clips.

8.6

Remove the ribbon from the Display PCB.

8.7

The front case is now removed.

9

Replacing the Front Case

9.1

Place the pump rear case (Fig 6, Item 20) on a flat surface with the inner components facing upward.

9.2

Plug the Main PCB - Display PCB ribbon cable into the rear of the display PCB ribbon connector (Fig 6, Item 50) within the front case.

9.3

Using thumbs carefully push down on the ribbon connector until the two securing clips ‘click’ into place over it.

9.4

Place the front case (Fig 6, Item 10) on top of the rear case (Fig 6, Item 20) ensuring that no wires are trapped.

9.5

Carefully turn the pump over so the front is face down. Install the five screws (Fig 6, Item 30) to the rear case and tighten.

Issue 1 April 2010

SER0018

Chap 5 Page 9

Pump Repair

Alpha Active 4 Service Manual

20 30

10

Figure 7 - Replacing Display PCB

Table 8 - Display PCB Parts List Item

Part Number

10

648420

20 30

Chap 5 Page 10

Description

Qty

Display PCB

1

REF

Screw (M3x8mm Pozi Pan Head, PT Type)

8

REF

Location Lug

-

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Pump Repair

10

Removing the Display PCB

10.1

Remove the Front case (Refer to Page 9, Section 8).

10.2

Remove the eight fixing screws (Fig 7, Item 20) from the display PCB.

10.3

Carefully lift the Display PCB (Fig 7, Item 10) out of the front casing.

11

Installing the Display PCB

11.1

Insert the base of the Display PCB (Fig 7, Item 10) into position within the front casing, guiding the square holes on the PCB to the location lugs (Fig 7, Item 30) within the front case.

11.2

Refit the eight fixing screws (Fig 7, Item 20).

11.3

Replace the Front case (Refer to Page 9, Section 9).

Issue 1 April 2010

SER0018

Chap 5 Page 11

Pump Repair

Alpha Active 4 Service Manual

10

20

30

50

40

Figure 8 - Compressor Fitment/Removal Table 9 - Compressor Assembly Item

Part Number

10

648408

20

Description

Qty

Compressor

1

REF

Compressor Outlet

-

30

REF

Compressor Mounting Plate

-

40

648425

Terminal Block

1

50

REF

T-Piece Connector

1

Chap 5 Page 12

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Pump Repair

12

Removing the Compressor Assembly

12.1

Remove the front case (Refer to Page 9, Section 8).

12.2

Remove the two red wires from the Terminal Block terminals T5 and T6.

12.3

Pull these wires free from other wiring within the rear case, noting the routing of the wires.

12.4

Remove the Compessor air outlet feed (Fig 8, Item 20) from the compressor (Fig 8, Item 10).

12.5

Lift the compressor assembly (Fig 8, Item 10) away.

13

Installing the Compressor Assembly

13.1

Place the compressor assembly into the rear case, ensuring the mounting plate (Fig 8, Item 30) sits in its moulding within the case.

13.2

Replace the compressor air feed (Fig 8, Item 20) to the new compressor (Fig 8, Item 10).

13.3

Route the compressor wires as required and connect to the appropriate terminals on the Terminal Block.

13.4

Reassemble the case (Refer to Page 9, Section 9).

Issue 1 April 2010

SER0018

Chap 5 Page 13

Pump Repair

Alpha Active 4 Service Manual

40

D C B

A

10

60

50

30

20

Microswitch omitted for clairity Figure 9 - Removing the Rotor/Stator Assembly Table 10 - Rotor/Stator Assembly Item

Part Number

10

648422

20

Description

Qty

Rotor/Stator Alpha Active 4

1

REF

Screw (M3x10mm Pozi Pan Head, PT Type)

2

30

REF

Rotor/Stator Spade Connectors

-

40

648427

Gearbox Assembly

1

50

648417

Microswitch

1

60

REF

Microswitch Securing Screw (M3x15mm Pozi Pan Head, PT Type)

2

Chap 5 Page 14

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Pump Repair

14

Removing the Microswitch Assembly

14.1

Remove the front case (Refer to Page 9, Section 8).

14.2

Remove the two spade connectors from the Rotor/Stator microswitch, noting wires/connections.

14.3

Remove the two securing screws (Fig 9, Item 60) that hold the microswitch in place.

14.4

You can now remove the Microswtich (Fig 9, Item 50).

15

Installing the Microswitch Assembly

15.1

Place the microswitch (Fig 9, Item 50) into position and secure with the two securing screws.

15.2

Re-connect the two spade connectors to the microswitch.

15.3

Re-fit the front case (Refer to Page 9, Section 9).

16

Removing the Rotor/Stator Assembly

16.1

Remove the front case (Refer to Page 9, Section 8).

16.2

Remove the four tubing connections to the Rotor/Stator, noting their connection locations.

16.3

Remove the two spade connectors from the Rotor/Stator microswitch, noting wires/connections.

16.4

Push down on the Rotor/Stator assembly Fig 9, Item 10, and remove the securing pin (D).

16.5

Remove the complete rotor/stator assembly (C,B,A) from the top of the gearbox.

17

Installing the Rotor/Stator Assembly

17.1

Refering to Figure 9: 17.1.1

Place the spring (A) onto the gearbox spindle as shown.

17.1.2

Place the bottom of the Rotor/Stator (B) onto the gearbox spindle.

17.1.3

Pushing down on the bottom section of the Rotor/Stator, place the top half of the Rotor/ Stator (C) onto the gearbox spindle AND while applying pressure, slide the securing pin (D) through the spindle as shown.

17.1.4

Ensure the securing pin (D) is seated within the recess on the top of the rotor/stator assembly.

17.2

Re-connect the two spade connectors to the Rotor/Stator microswitch.

17.3

Connect the four tubing connections to the Rotor/Stator.

17.4

Re-fit the front case (Refer to Page 9, Section 9).

18

Removing the Gearbox Assembly

18.1

Remove the rotor/stator assembly (Refer to Page 15, Section 16).

18.2

Remove the two securing screws (Fig 9, Item 20).

18.3

Remove the Gearbox-PCB connector from J3 on the Main PCB.

18.4

You can now remove the Gearbox Assembly (Fig 9, Item 40).

19

Installing the Gearbox Assembly

19.1

Place Gearbox assembly (Fig 9, Item 40) into position in the rear case, ensuring the wires exit the Gearbox in the direction of the compressor.

19.2

Secure with the two securing screws (Fig 9, Item 20).

19.3

Refer to “Installing the Rotor/Stator Assembly”.

Issue 1 April 2010

SER0018

Chap 5 Page 15

Pump Repair

Alpha Active 4 Service Manual

20

10

Figure 10 - Removing the Transformer Table 11 - Transformer Item

Part Number

10

648429

20

REF

Chap 5 Page 16

Description

Qty

Transformer Alpha Active 4

1

Screw (M3x10mm Pozi Pan Head, PT Type)

2

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Pump Repair

20

Removing the Transformer

20.1

Remove the front case (Refer to Page 9, Section 8).

20.2

On the main PCB, disconnect the transformer connection from J1.

20.3

Remove the red wires from Terminals T3 and T4 on the Terminal Block. Note: Ensure the capacitor remains connected to Terminals T3 and T4.

20.4

Remove the two securing screws (Fig 10, Item 20) that hold the transformer (Fig 10, Item 10) in place.

20.5

You can now remove the transformer (Fig 10, Item 10).

21

Installing the Transformer

21.1

Place the transformer (Fig 10, Item 10) into position in the rear case, ensuring it is correctly seated and orientation is so that the Main PCB wires are in the direction of the PCB.

21.2

Fit the two securing screws (Fig 10, Item 20).

21.3

Re-connect the red wires to Terminals T3 and T4 on the Terminal Block. Note: Ensure red wires run under all tubing, and that the capacitor remains connected to Terminals T3 and T4.

21.4

Re-connect the transformer connection to the Main PCB J1.

21.5

Re-fit the front case (Refer to Page 9, Section 9).

Issue 1 April 2010

SER0018

Chap 5 Page 17

Pump Repair

Alpha Active 4 Service Manual

30

A

20

10

Figure 11 - Removing the Main PCB Table 12 - Main PCB Parts List Item

Part Number

10

648421

20 30

Chap 5 Page 18

Description

Qty

Main PCB Alpha Active 4

1

REF

Stator To Pressure Switch Tubing

1

REF

Screw (M3x8mm Pozi Pan Head, PT Type)

4

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Pump Repair

22

Removing the Main PCB

22.1

Remove the front case (Refer to Page 9, Section 8).

22.2

On the Main PCB (Fig 11, Item 10), pull apart the two ribbon securing clips, at connector ‘A’.

22.3

Remove the ribbon cable from the Main PCB (Fig 11, Item 10).

22.4

On the main PCB, disconnect: 22.4.1

The transformer connection from J1.

22.4.2

The Rotor/Stator microswitch cable from J4.

22.4.3

The Rotor/Stator cable from J3.

22.4.4

The twin live connector from J7.

22.4.5

The 4-way connector from J2.

22.4.6

The Rotor/Stator tubing (Fig 11, Item 20) from U3.

22.5

Remove the four corner fixing screws (Fig 11, Item 30) and lift the Main PCB (Fig 11, Item 10) out of the pump case.

23

Replacing the Main PCB

23.1

Place the Main PCB with its component side facing upward, into position in the rear case, ensuring alignment of the four corner fixing holes in Main PCB (Fig 11, Item 10) with the pillars in the rear case. Fit the four retaining screws (Fig 11, Item 30).

23.2

Connect to the Main PCB:

23.3

23.2.1

The Rotor/Stator tubing to U3.

23.2.2

The 4-way connector to J2.

23.2.3

The twin live connector to J7.

23.2.4

The Rotor/Stator cable to J3.

23.2.5

The Rotor/Stator microswitch cable to J4.

23.2.6

The transformer connection to J1.

Plug the ribbon cable into the connector on the Main PCB within the rear case. Note: If Main PCB - Display PCB ribbon cable has been disconnected from both the PCB’s, ensure upon re-connecting that the location keys are in correct orientation.

23.4

Using thumbs, carefully push down on the ribbon connector until the two securing clips ‘click’ into place over it.

23.5

Re-fit the front case (Refer to Page 9, Section 9).

Issue 1 April 2010

SER0018

Chap 5 Page 19

Pump Repair

Alpha Active 4 Service Manual

20

30

10

Figure 12 - Removing the Power Fail Battery Table 13 - Power Fail Battery Parts List Item

Part Number

10

648421

Main PCB Alpha Active 4

1

20

648428

Battery

1

30

BP196

Cable Tie

1

Chap 5 Page 20

Description

SER0018

Qty

Issue 1 April 2010

Alpha Active 4 Service Manual

Pump Repair

24

Removing Power Fail Battery

24.1

Remove the front case (Refer to Page 9, Section 8).

24.2

Remove the Main PCB (Refer to Page 19, Section 22).

24.3

De-solder the battery connections from the PCB.

25

Replacing Power Fail Battery

25.1

Place the battery (Fig 12, Item 20) onto the PCB (Fig 12, Item 10) ensuring the correct polarity.

25.2

Solder both the positive and negative connections.

25.3

Secure battery to PCB using tie wrap (Fig 12, Item 30).

25.4

Replace the PCB (Refer to Page 19, Section 23).

25.5

Re-fit the front case (Refer to Page 9, Section 9).

Issue 1 April 2010

SER0018

Chap 5 Page 21

Pump Repair

Alpha Active 4 Service Manual

Rear of On/Off Switch 10

Figure 13 - Removing the On/Off Switch Table 14 - On/Off Switch Parts Item

Part Number

10

648407

Chap 5 Page 22

Description Mains (On/Off) Switch

SER0018

Qty 1

Issue 1 April 2010

Alpha Active 4 Service Manual

Pump Repair

26

Removing the On/Off Switch

26.1

Remove the front case (Refer to Page 9, Section 8).

26.2

Lift the On/Off switch (Fig 13, Item 10) away from the rear pump case.

26.3

Remove the two live spade connectors from the back of the On/Off Switch (terminals 1 and 1A).

26.4

Remove the two neutral spade connectors from the back of the On/Off Switch (terminals 2 and 2A).

26.5

You can now remove the On/Off Switch (Fig 13, Item 10).

27

Installing the On/Off Switch

27.1

Re-connect the live spade connector that runs from the Main PCB to terminal 1 (Refer to Figure 13)on the rear of the On/Off Switch.

27.2

Re-connect the live spade connector that runs from a fuse holder to terminal 1a (Refer to Figure 13) on the rear of the On/Off Switch.

27.3

Re-connect the neutral spade connector that runs from the Terminal block and Main PCB to spade terminal 2 (Refer to Figure 13) on the rear of the On/Off Switch.

27.4

Re-connect the neutral spade connector that runs from a fuse holder to spade terminal 2a (Refer to Figure 13) on the rear of the On/Off Switch.

27.5

Place the On/Off Switch (Fig 13, Item 10) into position in the rear case, ensuring the two live wires are facing outwards from the case.

27.6

Re-fit the front case (Refer to Page 9, Section 9).

Issue 1 April 2010

SER0018

Chap 5 Page 23

Pump Repair

Alpha Active 4 Service Manual

10

Figure 14 - Removing the Fuse Holder Housings Table 15 - Fuse Holder Housings Item

Part Number

10

648406

Chap 5 Page 24

Description Fuse Holder

Qty 2

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Pump Repair

28

Removing the Fuse Holder Housing

28.1

Remove the front case (Refer to Page 9, Section 8).

28.2

For the right Fuse Holder Housing:

28.3

28.2.1

Remove the live Mains/Power Cable spade connector from the side of the fuse holder.

28.2.2

Remove the live On/Off Switch spade connector from the top of the fuse holder.

28.2.3

Unscrew nut and remove the Fuse Holder from the pump.

For the left Fuse Holder Housing: 28.3.4

Remove the neutral Mains/Power Cable spade connector from the side of the fuse holder.

28.3.5

Remove the neutral On/Off Switch spade connector from the top of the fuse holder.

28.3.6

Unscrew nut and remove the Fuse Holder from the pump.

29

Installing the Fuse Holder Housing

29.1

For the right Fuse Holder Housing:

29.2

29.3

29.1.1

Secure the Fuse Holder to the pump.

29.1.2

Connect the live On/Off Switch spade connector to the top of the fuse holder.

29.1.3

Connect the live Mains/Power Cable spade connector to the side of the fuse holder.

For the left Fuse Holder Housing: 29.2.4

Secure the Fuse Holder to the pump.

29.2.5

Connect the neutral On/Off Switch spade connector to the top of the fuse holder.

29.2.6

Connect the neutral Mains/Power Cable spade connector to the side of the fuse holder.

Re-fit the front case (Refer to Page 9, Section 9).

Issue 1 April 2010

SER0018

Chap 5 Page 25

Pump Repair

Alpha Active 4 Service Manual

10

20

30

Figure 15 - Removing the Mains/Power Cable Table 16 - Mains/Power Cable Item

Part Number

10

648410

Mains Cable Euro - Alpha Active 4

1

20

648447

Ferrite 28mm

1

30

648449

Ferrite 12mm

1

Chap 5 Page 26

Description

SER0018

Qty

Issue 1 April 2010

Alpha Active 4 Service Manual

Pump Repair

30

Removing the Mains/Power Cable

30.1

Remove the front case (Refer to Page 9, Section 8).

30.2

Remove the live spade connection from the side of the right fuse holder.

30.3

Remove the neutral spade connection from the side of the left fuse holder.

30.4

Remove the two ferrite beads from the Mains/Power Cable.

30.5

You can now remove the Mains/Power Cable.

31

Installing the Mains/Power Cable

31.1

Place the Mains/Power Cable in position in the side of the rear case.

31.2

Re-connect the neutral spade connector to the side of the left fuse holder.

31.3

Re-connect the live spade connector to the side of the right fuse holder.

31.4

Re-fit the front case (Refer to Page 9, Section 9).

Issue 1 April 2010

SER0018

Chap 5 Page 27

Pump Repair

Alpha Active 4 Service Manual

30

30

20 10

Figure 16 - Removing the Terminal Block Table 17 - Terminal Block Item

Part Number

10

648425

20

REF

30

648449

Chap 5 Page 28

Description

Qty

Terminal Block Alpha Active 4

1

Screw (M3x15mm Pozi Pan Head, PT Type)

1

Filter Capacitor

1

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Pump Repair

32

Removing the Terminal Block

32.1

Remove the front case (Refer to Page 9, Section 8).

32.2

Remove the red Transformer wires and the capacitor from Terminals T3 and T4 on the Terminal Block (Fig 16, Item 10).

32.3

Remove the red Main PCB wires from Terminals T1 and T2 on the Terminal Block (Fig 16, Item 10).

32.4

Remove the two live wires from Terminal T8 on the Terminal Block (Fig 16, Item 10).

32.5

Remove the neutral wire from Terminal T7 on the Terminal Block (Fig 16, Item 10).

32.6

Remove the two red compressor wires from Terminals T5 and T6 on the Terminal Block (Fig 16, Item 10).

32.7

Remove the single securing screw (Fig 16, Item 20).

32.8

You can now remove the Terminal Block (Fig 16, Item 10).

33

Installing the Terminal Block

CAUTION: Whilst completing the installation, ensure all wires are free from entrapment. 33.1

Place the Terminal Block (Fig 16, Item 10) into position ensuring it is seated correctly.

33.2

Replace the single securing screw (Fig 16, Item 20).

33.3

Replace the two red compressor wires to Terminals T5 and T6 on the Terminal Block (Fig 16, Item 10).

33.4

Replace the neutral wire from the Main PCB Connector J2 and the On/Off switch to Terminal T7 on the Terminal Block (Fig 16, Item 10).

33.5

Replace the two live wires from the Main PCB Connector J2 and the On/Off switch to Terminal T8 on the Terminal Block (Fig 16, Item 10).

33.6

Replace the red Transformer wires, and the capacitor to Terminals T3 and T4 on the Terminal Block (Fig 16, Item 10).

33.7

Replace the red Main PCB wires to Terminals T1 and T2 on the Terminal Block (Fig 16, Item 10).

33.8

Re-fit the front case (Refer to Page 9, Section 9).

Issue 1 April 2010

SER0018

Chap 5 Page 29

Pump Repair

Chap 5 Page 30

Alpha Active 4 Service Manual

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Mattress Repair

CHAPTER 6 MATTRESS REPAIR 1

General

This chapter details repair procedures for the Alpha Active 4 Mattress Replacement. All repairs should be carried out by ArjoHuntleigh approved service personnel. It is easier to carry out the repair procedures in this chapter if the mattress is fully deflated first. Disconnect the tubeset between the mattress and the pump to deflate the mattress. After carrying out a service or any of these repairs on the mattress, check the following: 1.1

All press studs on the mattress are connected.

1.2

All cell assemblies are securely held by their straps. The repair will normally have disturbed the position of the other cells, and they should be re-positioned accordingly.

2

Alpha Active 4 Cell Layout

Figure 17 - Mattress Cell Layout

Issue 1 April 2010

SER0018

Chap 6 Page 1

Mattress Repair

Alpha Active 4 Service Manual

10

50

40

30 20

Figure 18 - Alpha Active 4 Mattress Table 18 - Alpha Active 4 Mattress Item

Part Number

10

648437

Mattress Top Cover (90cm Wide)

1

-

648461

Mattress Top Cover (85cm Wide)

-

20

648438

Mattress Base Cover (90cm Wide)

1

648463

Mattress Base Cover (85cm Wide)

648465

Mattress Base Cover PU (85cmWide)

30

REF

40 50

Chap 6 Page 2

Description

Qty

Zip

2

648442

Tubeset Connector - Alpha Active 4

1

648443

Tri bore tubing. 9.5mm ID x 14.5mm OD x 1370mm

1

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Mattress Repair

3

Removing The Top Cover

3.1

Fully unroll the mattress on a suitable surface.

3.2

Undo the zips (Fig 18, Item 30) on both sides of the top cover.

3.3

Remove the top cover.

4

Installing the Top Cover

4.1

Position the top cover (Fig 18, Item 10) on to the mattress. Make sure the feet symbol is at the foot end of the mattress.

4.2

Secure the two zips on the long sides of the mattress (Fig 18, Item 30) to the correct side of the top cover (Fig 18, Item 10).

Issue 1 April 2010

SER0018

Chap 6 Page 3

Mattress Repair

Alpha Active 4 Service Manual

10

60

50

30

Cell

20 40

Feed To Head End

20

30

Feed From Foot End

Figure 19 - Cell Assembly

Chap 6 Page 4

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Mattress Repair

5

Removing Cell Assembly

5.1

Remove the top cover, (Refer to Page 3, Section 3).

5.2

Disconnect the press studs from each end of the cell ((Fig 19, Item 60).

5.3

Disconnect the lower (Fig 19, Item 30) and upper (Fig 19, Item 20) cell feed connectors from the T-Pieces (Fig 19, Item 40).

5.4

Pull the cell out of the retaining loop in the base cover and remove from the mattress.

6

Installing Cell Assembly

6.1

Pass the cell through the loop on the air pad cover and centralise the cell.

6.2

Connect the upper (Fig 19, Item 20) and lower (Fig 19, Item 30) cell feed connectors to each TPieces (Fig 19, Item 40).

Note: Installing the cell feeds is made easier if the cell connector and/or T-Piece is wetted immediately prior to insertion. 6.3

Reconnect the male press studs to the female studs (Fig 19, Item 60) at the end of the cell.

6.4

Install the top cover (Refer to Page 3, Section 4).

Table 19 - Cell Assembly Parts List Item

Part Number

Description

Qty

10

648439

Standard Cell (90cm Wide)

11

648440

Ventilated Cell (90cm Wide)

8

648467

Standard Cell (85cm Wide)

648468

Ventilated Cell (85cm Wide)

648470

Standard Cell PU(85cm Wide)

648471

Ventilated Cell PU (85cm Wide)

20

REF

Upper Cell Feed (Part of Item 10)

-

30

REF

Lower (Static) Cell Feed (Part of Item 10)

-

40

REF

T-Piece

50

REF

Single Bore Tubing

-

-

REF

Press Stud Male

1

60

REF

Press Stud Female (Located Inside Mattress Base Cover)

1

Issue 1 April 2010

38

SER0018

Chap 6 Page 5

Mattress Repair

Alpha Active 4 Service Manual

40 A 20 30 20

90

30

50 60

30

30

80

20

50

20 80

20

20 20

20 60 20

20 60

20 20 60 20 20

20

60 20 60

20 20 10 80

60

70

20 B - Continued from point ‘A’ Figure 20 - Manifold Tubing

Chap 6 Page 6

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Mattress Repair

7

Replacing Manifold Tubing

7.1

Remove the top cover, (Refer to Page 3, Section 3).

7.2

Disconnect the press studs from each end of the cell ((Fig 19, Item 60).

7.3

Disconnect the lower (Fig 19, Item 30) and upper (Fig 19, Item 20) cell feed connectors from the T-Pieces (Fig 19, Item 40).

7.4

Connect the upper (Fig 19, Item 20) and lower (Fig 19, Item 30) cell feed connectors to each TPieces (Fig 19, Item 40).

Note: Installing the cell feeds is made easier if the cell connector and/or T-Piece is wetted immediately prior to insertion. 7.5

Reconnect the male press studs to the female studs (Fig 19, Item 60) at the end of the cell.

7.6

Install the top cover (Refer to Page 3, Section 4).

Table 20 - Cell Assembly Parts List Item

Part Number

10

REF

50mm (Cut to length from 648436)

20

REF

84mm (Cut to length from 648436)

30

REF

40mm (Cut to length from 648436)

40

REF

140mm (Cut to length from 648436)

50

REF

64mm (Cut to length from 648436)

60

REF

184mm (Cut to length from 648436)

-

648436

Single bore tubing: ID 9.5mm x OD14.5mm

70

648443

Tri Bore Tubing

80

648444

Static Check Valve

3

90

648441

CPR

1

Issue 1 April 2010

Description

SER0018

Qty

Chap 6 Page 7

Mattress Repair

Alpha Active 4 Service Manual

10

Head Cells Removed For Drawing Clarification

Figure 21 - CPR Assembly Table 21 - CPR Assembly Parts List Item

Part Number

10

648441

Chap 6 Page 8

Description CPR

Qty 1

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Mattress Repair

8

Removing CPR Assembly

8.1

Remove the top cover, (Refer to Page 3, Section 3).

8.2

Disconnect the press studs from cells either side of the CPR assembly.

8.3

Disconnect the feed tubes from either side of the CPR unit.

8.4

Remove CPR Assembly from mattress (Fig 21, Item 10).

9

Installing CPR Assembly

9.1

Connect the feed tubes from either side of the CPR unit.

Note: Installing the feed tubes is made easier if the CPR unit connectors are wetted immediately prior to insertion. 9.2

Reconnect the press studs to each end of the cell.

9.3

Install the top cover (Refer to Page 3, Section 4).

Issue 1 April 2010

SER0018

Chap 6 Page 9

Mattress Repair

Alpha Active 4 Service Manual

10 Red

Yellow

To Cells

To Pump

20

Figure 22 - Static Feed Check Valve Table 22 - Static Section Feed Valve Parts List Item

Part Number

10

648444

Static Feed Check Valve

3

20

BP196

Tie Wrap

6

Chap 6 Page 10

Description

SER0018

Qty

Issue 1 April 2010

Alpha Active 4 Service Manual

10

Mattress Repair

Removing Static Feed Check Valves

Note: There are 3 Static Feed Check Valves on the Alpha Active 4. Two at the head end, and one at the foot end. 10.1

Remove the top cover, (Refer to Page 3, Section 3).

10.2

Release the pop studs on three surrounding cells for easier access to the Check Valve.

10.3

Remove the two cable ties (Fig 22, Item 20) located on either side of the Check Valve (Fig 22, Item 10) barbs.

10.4

Seperate the tubes away from the Check Valve.

10.5

Remove the Check Valve (Fig 22, Item 10).

11

Installing Static Feed Check Valves

CAUTION: Ensure correct orientation is followed during this installation (Refer to Page 10, Figure 22). 11.1

Place the Check Valve (Fig 22, Item 10) in position.

11.2

Connect the two tubes to the Check Valve (Fig 22, Item 10) barbs.

11.3

Secure the tubing with two cable ties (Fig 22, Item 20).

11.4

Secure the three cells pop studs that were released in Para 10.2.

11.5

Install the top cover (Refer to Page 3, Section 4).

Issue 1 April 2010

SER0018

Chap 6 Page 11

Mattress Repair

Alpha Active 4 Service Manual

20

10

10 20

10

Figure 23 - 3-Way Connector Table 23 - 3-Way Connector Parts List Item

Part Number

10

648442

3-Way Connector

1

20

648443

Tri bore tubing. 9.5mm ID x 14.5mm OD x 1370mm

1

Chap 6 Page 12

Description

SER0018

Qty

Issue 1 April 2010

Alpha Active 4 Service Manual

Mattress Repair

12

Removing 3-Way Connector

12.1

With the tubeset (Fig 23, Item 20) held securely in one hand, pull the top of the 3-way connector (Fig 23, Item 10) away from the tubeset as shown in Figure 23 1.

12.2

Pull the bottom of the 3-way connector away from the tubeset (Figure 23 2).

12.3

You can now pull the 3-way connector away 3 (Fig 23, Item 10).

13

Installing 3-Way Connector

Note: Orientation of this part is un-important. 13.1

Align the tubeset (Fig 23, Item 20) with the barbs on the rear of the 3-way connector (Fig 23, Item 10).

13.2

Push the tubeset fully onto the connector until all three tubes contact the base of the connector.

13.3

Ensure the 3-way connector (Fig 23, Item 10) is securely fitted to the tubeset (Fig 23, Item 20).

Issue 1 April 2010

SER0018

Chap 6 Page 13

Mattress Repair

Chap 6 Page 14

Alpha Active 4 Service Manual

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Technical Specification

CHAPTER 7 TECHNICAL SPECIFICATION PUMP Model:

Alpha Active 4

Part Numbers:

648311

UK

648314

German

648315

Euro

648319AU

Australia

Supply Voltage:

230V

Supply Frequency:

50-60Hz

Power Input:

230VA

Size:

(L)280mm x (W)205mm x (H)112mm

Weight:

2.5kg

Case Material:

ABS Plastic

Plug Fuse Rating:

5A to BS1362 (UK ONLY)

Pump Fuse Rating:

T1AL 250V

Degree of protection against electric shock:

Class II, Double Insulated Type BF

Degree of protection against liquid ingress:

IPX0

Mode of operation:

Continuous

Cycle Times:

10 mins

20 mins

Inflate - 4.5 mins

Inflate - 9.5 mins

Crossover - 30 secs

Crossover - 30 secs

Deflate - 4.5 mins

Deflate - 9.5 mins

Crossover - 30 secs

Crossover - 30 secs

SYMBOLS O (Off)

Power Disconnects from the mains supply

I (On)

Power Connects to the mains supply Dangerous voltage

Issue 1 April 2010

i

Do not dispose of in the domestic refuse

Alternating Current

Double Insulated

Refer to accompanying documents

Type BF

Refer to the User Manual

Fuse (If shown on pump label)

Ref:

SER0018

Model number

SN:

Serial Number

Chap 7 Page 1

Technical Specification

Alpha Active 4 Service Manual

ENVIRONMENTAL INFORMATION Condition

Temperature Range

Relative Humidity

Atmospheric Pressure

Operating (Pump)

+10°C to +40°C

30% to 75%

700hPa to 1060 hPa

Storage and Transport (Pump)

-40°C to +70°C

10% to 95% (non-condensing)

500 hPa to 1060 hPa

MATTRESS Description

Cell Material

Base Pad Material

ALPHA ACTIVE 4 MR 90

Polyurethane

PVC Nylon Coated

ALPHA ACTIVE 4 MR 85

Polyurethane

PVC Nylon Coated

ALPHA ACTIVE 4 MR Non PVC 85

Polyurethane

PU Coated

MATTRESS SIZE INFORMATION Part No.

Description

Spare Cover

Length mm

Width mm

Height mm

648322

ALPHA ACTIVE 4 MR 90

648437

1950 (77")

900(35")

200 (8")

648324

ALPHA ACTIVE 4 MR 85

648461

1950 (77")

850(33")

200 (8")

648326

ALPHA ACTIVE 4 MR 85 PU

648461

1950 (77")

850(33")

200 (8")

Chap 7 Page 2

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Technical Specification

COVER SPECIFICATION Standard Cover (Dartex)®

Feature Removable Cover

Yes

Moisture Vapour Permeable

Yes

Air Permeable

No

Low Friction

Yes

Water Resistant / Repellent

Yes Material coating is Bacteriostatic, fungistatic, antimicrobial

Infection Control Fire Retardant

BS 7175: 0,1 & 5

2-Way Stretch

Yes MAX 95°C (203°F) for 15 mins(1)

Washing Conditions

Tumble Dry up to 130°C (266°F) or Air Dry

Drying Conditions

50 Wash Cycles (minimum)

Life Span

Acute and Homecare

Application Area

1. Check your local policy to determine the time/temperature ratio required to achieve thermal disinfection.

CLEANING SYMBOLS

Issue 1 April 2010

Wash at 80°C (176°F)

Do not tumble dry above 50°C.

Do not iron

Do Not Use Phenol-based cleaning Solutions

Wipe surface with damp cloth

Use solution diluted to 1000 ppm of Available Chlorine

SER0018

Chap 7 Page 3

Technical Specification

Chap 7 Page 4

Alpha Active 4 Service Manual

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Spare Parts

CHAPTER 8 SPARE PARTS 1

Pump Assembly Parts List

Part Number

Description

Fig - Item

Qty

648410

• Mains/Power Cable - Euro

15-10

1

648418

• Mains/Power Cable - UK

-

-

648419

• Mains/Power Cable - AUS

-

-

648447

• Ferite 28mm

15-20

1

648449

• Ferite 12mm

15-30

1

648401

• Front Case

6-10

1

648402

• Rear Case

3-10

1

648403

• Bed Hook

4-30

2

648404

• Air Filter

3-NA

1

648405

• Air Filter Grill

3-20

1

648420

• Display PCB

6-40

1

648421

• Main PCB

11-10

1

648428

• Battery

12-20

1

648425

• Terminal Block

16-10

1

648408

• Compressor

8-10

1

648422

• Rotor/Stator Assembly

9-10

1

648429

• Transformer

10-10

1

648449

• Filter Capacitor

16-30

1

MIS106

• Mains Fuse (1A Time Lag T Fuse)

5-20

2

648423

• Hose Connector Alpha Active 4

6-60

1

648406

• Fuse Holder

14-10

2

648407

• Mains Switich

13-10

1

2

Mattress Assembly Parts List

Part Number

Description

Fig - Item

Qty

648437

• Top Cover (195cm x 90cm) - printed with logo

18-10

1

648461

• Top Cover (195cm x 85cm) - printed with logo

-

-

648438

• Base Cover (195cm x 90cm)

18-20

1

Issue 1 April 2010

SER0018

Chap 8 Page 1

Spare Parts

Alpha Active 4 Service Manual

Part Number

Description

Fig - Item

Qty

648463

• Base Cover (195cm x 85cm)

-

-

648465

• Base Cover PU (195cm x 85cm)

-

-

648442

• Tubeset Connector Alpha Active 4

23-10

1

648444

• Static Feed Check Valve

22-10

3

648441

• CPR Assembly

21-10

1

648439

• Standard Cell (90cm Wide)

19-10

11

648440

• Ventilated Cell (90cm Wide)

19-10

8

648467

• Standard Cell (85cm Wide)

-

-

648468

• Ventilated Cell (85cm Wide)

-

-

648470

• Standard Cell PU (85cm Wide)

-

-

648471

• Ventilated Cell PU (85cm Wide)

-

-

648436

• Single bore tubing

20-NA

-

648443

• Tri bore tubing

20-70

-

20-21

6

• BP196

Chap 8 Page 2

• Tie Wrap

SER0018

Issue 1 April 2010

Alpha Active 4 Service Manual

Service Contact Details

CHAPTER 9 SERVICE CONTACT DETAILS Africa

Australia

Huntleigh Africa (Pty) Ltd William Cruywagen Avenue Stand 120, Klerksoord, Pretoria T +27 12 542 4680 F +27 12 542 4982

ArjoHuntleigh Pty Limited PO Box 330 Hamilton Hill Western Australia 6963 T +61 8 9 337 4111 F +61 8 9 337 9077

Belgium

Denmark

ArjoHuntleigh NV SA Stapelplein 70, 9000 Gent, Belgium T +32 9 265 8770 F +32 9 265 8771

ArjoHuntleigh A/S Vassingerødvej 52, 3540 Lynge, Denmark T +45 4 913 8486 F +45 4 913 8487

France

Germany

HNE Médical 451 Chemin de Champivost BP20, 69579 Limonest, Cedex, France T +33 (0)4 78 66 62 66 F +33 (0)4 78 66 62 67

HNE Huntleigh Nesbit Evans Healthcare GmbH Im Hülsenfeld 19, 40721 Hilden, Germany T +49 (0)2103 97 11 0 F +49 (0)2103 97 11 80

Holland

India

ArjoHuntleigh BV Antennestraat 45, 1322 AH Almere, Holland T +31 36 533 5588 F +31 36 547 5075

ArjoHuntleigh India Pvt. Ltd 8, Shah Industrial estate Off, Shah Industrial estate Off Veera Desai Road Andheri (West) Mumbai – 400 053 India

Middle East

New Zealand

Huntleigh Technology PLC 307 Sh. Zayed Road Dubai United Arab Emirates

ArjoHuntleigh Ltd Unit 6/38 Eaglehurst Road Ellerslie, Auckland, NZ T +64 9 525 2488 F +64 9 525 2433

Singapore

Sweden

ArjoHuntleigh Pte Limited Block 603, Elias Road, #08-232 Singapore 510603 T +61 8 9 309 3083 F +60 8 9 309 4582

Huntleigh Liljenberg AB Box 30012 200 61 Limhamm, Sweden T +46 40 36 03 50 F +46 40 49 43 75

Issue 1 April 2010

SER0018

Chap 9 Page 1

Service Contact Details

Alpha Active 4 Service Manual

UK

United States of America

ArjoHuntleigh Limited 310-312 Dallow Road Luton Bedfordshire LU1 1TD T +44 (0)1582 413104 F +44 (0)1582 459100

ArjoHuntleigh 2349 W Lake Street- Suite 250 Addison, IL 60101, USA Tel: +1 630 307 2756 Toll Free US: (800) 323 1245 Fax: +1 630 307 6195

Spain

Japan

ArjoHuntleigh SL C/ de la Creueta 25 08130-Santa Perpetua de la Mogoda, Barcelona, T +34 93 5745754 F +34 93 5745861

ArjoHuntleigh Japan K.K. 4-1-15 Goko-Dori Chuoku Kobe 651-0087 T +81 78 231 8735 F +81 78 231 8736

Chap 9 Page 2

SER0018

Issue 1 April 2010

SER0018_01 April 2010