and a Agilent 6210 MSD TOF mass spectrometer was used (Figure 2). Experimental ... LC/TOF instrument configuration. Figure 1. Topology Of Silica B And ...
The The Use Use Of Of A A Novel Novel Type-C Type-C Silica Silica To To Separate Separate Amino Amino Acids Acids By By Aqueous Aqueous Normal Phase (ANP) With Electrospray (ESI) Detection Normal Phase (ANP) With Electrospray (ESI) Detection
An Agilent 1100 Series HPLC system with binary pump and degasser, well plate autosampler with thermostat, thermostatted column compartment, and a Agilent 6210 MSD TOF mass spectrometer was used (Figure 2).
Amino Acid Retention Data Retention time data was collected under isocratic conditions of 50%, 60%, 70%, 75%, 80%, 85% and 90% organic solvent. The organic solvent was composed of 0.1% formic acid in acetonitrile (Figure 3), methanol (Figure 4) or acetone (Figure 5). Water containing 0.1% formic acid made up the difference. The mobile phase flow rate was 0.4 ml/min. The column temperature was 20oC. The column t0 is 1.44 minutes.
Figures 3 -5 are the retention plots for the amino acids investigated in this study. For acetonitrile and acetone there is a substantial increase in retention as the percentage of organic in the mobile phase exceeds 60%. For methanol only a small increase in retention is observed. Thus it appears that aprotic solvents induce greater ANP behavior than protic solvents. Use of acetone which appears to have good ANP properties is possible in LC/MS but not conventional UV detection. Another advantage of the use of ANP is the increase in sensitivity at high organic content in the mobile phase. Typical additives needed to promote ANP retention (acetic or formic acid) are very compatible with mass spectroscopy detection.
50
4 3 2
0 50
60
70
0.25
Retention time reproducibility data was collected for one gradient at 15 and 30oC. The flow rate was 0.4 ml/min and an equilibration time of 5 minutes. The gradient was;
0.15
Time
Retention time vs % acetone in the mobile phase
1
40
Gradient Reproducibility
80
90
100
% Acetone
%B
0
90
5 6
Time
0.2
%B
8
80
90
9
50
80
13
50
Table 2 shows retention time reproducibility for nine amino acids at two temperatures. Four replicates were performed at each temperature. The reproducibility was 0.28% or better for the amino acids. This is a significant improvement over what is usually observed for most HILIC analyses, especially considering this is gradient data with only a 5 minute re-equilibration time between runs.
0.1 0.05 10.5
Lys
Individual amino acid standards were purchased from Sigma-Aldrich. The amino acids studied are in Table 1.
Figure 7 shows the extracted ion chromatogram (EIC) of a nineteen amino acid mixture under one of the gradient conditions tested. All of the critical amino acid pairs (those that are isobaric or have masses within one mass unit ) are separated under these conditions except for the Leucine / Isoleucine pair. At present, the maximum separation is approximately 0.15 min with 0.30 minutes needed for resolution with the peak widths obtained for these two amino acids. Additional gradient formats and mobile phases are under investigation to address this issue.
Figure 6 shows the retention behavior of amino acids as a function of temperature (log tR vs. 1/T, vant Hoff plot). In all cases, retention increases as the temperature is increased indicating either a positive enthalpy for interaction of the solute with the stationary phase or substantial entropy contributions (proton activity). This temperature effect is opposite of what is typically observed under reverse phase conditions i.e. decreasing retention with increasing temperature.
Retention time vs. % acetonitrile in the mobile phase
Results and Discussion
Val
A high surface area, 4.0 µm particle size Type-B silica was converted to Type-C material having a silica hydride (Si-H) surface (Figure 1). To improve peak shape of the basic amino acids the Type-C silica was further reacted to reduce residual silanols (
A CD with the Excel-based Template program, along with electronic copies of the manuals and ... For manual flow rate qualification, a typical range for an analytical HPLC ..... enable the Macros, by setting the security level to Low or Medium.
Description. APP A44. APPLICA. TION. NOTES. Method Conditions. Column: ... retention higher percentages of water in the mobile phase can be used. The diminished hydrophobicity is shown in Table 1. For more information visit www.
40018-75P. Cogent Bidentate⢠C18 HPLC Column, 4µm \,100A 4.6mm x 75mm. Cat. ... Solvent A: 100% DI water + ... 1mg of the compound dissolved in 1 mL.
tion There is no need to filter any of the CElixir Solution (A) before coating ... celerator Solution (B) can be used as your sample diluent but you must be careful to ...
Welcome to the PQ Kit! These instructions are designed to help you to quickly familiarize yourself with the procedures needed to fully qualify your HPLC using ...
Mass Spec: Atmospheric Pressure. Chemical Ionization in positive mode: APCI+ Single Ion Monitoring. Discussion. Two proprietary compounds, which are ...
ESI â pos - Agilent 6210 MSD TOF mass spectrometer. Discussion ... can be used for accurate determination of NT and metabolites in biological samples for ...
LC-MS method for analysis of trace levels of melamine. Excellent ... ESI â pos - Agilent 6210 MSD TOF MS. ... when using LC-MS on standard HPLC columns.
(5 mg/mL (free base) of each dissolved in the mobile phase). Detection: Mass Spec: Atmospheric Pressure. Chemical Ionization in positive mode: APCI+ Single ...
Uracil (m/z 113), Tobramycin (m/z 468). 1 mg/L of each dissolved in DI water. Detection: Mass Spec: Atmospheric Pressure. Chemical Ionization in positive mode ...
For these compounds, CElixir-SDS should be considered as an alter- .... Methanol, Ethanol, Acetonitrile, Isopropanol Methoxyethanol, Ethylene Glycol. Run.
pH changing when moving to the acidic range. Start with the CElixir⢠Accelerator Solution (B) pH 2.5. CElixirâ¢. Accelerator Solution is used as your run buffer ...
To Order Call: 1-732-578-1777 or Order online @ www.mtc-usa.com. 2. For Most ... Terms are net 30 days for cus- .... These center .... Q Analytical Glossary.
use pH as a solubility tool, a selectivity tool or as a sample matrix option to retain the biological activity of your ... Save Money buying solvents and disposal costs.
tures (such as long chain or multi ring hydrocarbons) and small for compounds with highly polar groups. Log P values provides a measure of hydrophilic vs.
Vials# 29, 31 â CE grade Water. 4. Vial# 32 â CElixir ... Stop Collecting data at the end of your run, for example 15.0 minutes. Rinse. 10. ... 11. Using pressure, run CEwater through the capillary using Inlet Vial#29 and Outlet. Vial# 10 for 0.5
Cogent Bidentate C18 HPLC Column, 100à , 4µm, 4.6 x 75 mm, Standard End Fittings. Cat. No. ... refraction indices between solvent A and solvent B in gradient ...
69069-Q20. Cogent Qx2⢠Column for LC-MS, 2.1 ID x 20 mm, PEEK & Titanium housing, ... Mass Spec: Atmospheric Pressure. Chemical Ionization in positive.
By using ANP for analysis of polar compounds by LC-MS a 10-100 fold increase of sensitivity is ... dissolved in water or water + organic solvent mixture, which is ...
Cogent Diamond Hydrideâ¢HPLC Column, 100à , 4µm, 2.1 x150 mm. Cat. No. Description ... diseases: congestive heart failure or dehydration. It may also be ...